
ENERGY, ENVIRONMENTAL &
CHEMICAL ENGINEERING
Molecular and Nanoscale Analysis Laboratory
ICP-MS Inductively Coupled Plasma Spectrometer
(Agilent 7500 ce)
Instrument for Routine Elemental Analysis
Introduction
Inductively Coupled Plasma is commonly abbreviated as ICP. It is a fast,
high
sensitivity, high resolution instrument that is capable of of multi-element analysis
simultaneously.

Structure of the ICP
The major components of
the instrument are:
Power Supply: mainly provides the energy for the argon plasma.
Argon Gas: provides argon gas to the plasma and cooling
gas to the torch.
Cooling System: removes any excess amount of heat from the system, specially the RF
system.
Sampling System: includes the pump and aerosol to feed
the system and autosampler.
Optical Unit: distinguishes the signals from different
wavelengths, amplify the signal and
covert an optical signal to an electrical signal.
Standard
Operation Procedure (SOP) Of ICP
Turn on ICP
- Turn gas supply on at
the cylinder (green knob tip points towards hose)
- Make sure the locking
lever on the plasma compartment door is securely latched.
- Switch PC Monitor on
- Switch on Water Cooler
- Turn on the power to the
instrument (front panel I/O switch to green light on)
Wavelength
Calibration
Note:
It is recommended that the system be allowed to warm up for at least 15 minutes
before the wavelength calibration is carried out.
A wavelength calibration determines the
grating position/wavelength relationship.
You must perform a wavelength calibration before the instrument can be used.
The grating is calibrated using a number of Hg lines in various diffraction
orders.
- Go to the
Wavelength Calibration page of the
Instrument Setup window (click ‘Instrument’ icon on start-up page).
- Click
Calibrate.
The instrument will switch on the Hg lamp and
allow it to warm up before starting the calibration.
The grating is calibrated
using eleven Hg lines to determine the grating position/wavelength
relationship. When the calibration is complete, the instrument is ready for the
plasma to be
ignited.
Ignite Plasma
- Put sample tubing in DI water.
- Click “Plasma On”.
- Gas pressure should be at 70 pis.
- Wait 30 minutes for stabilization.
Performing a torch
alignment scan
-
Place the sample pump tubing from
the peristaltic pump into a 5 ppm Mn solution.
-
Click Instrument Setup on the Window menu. The
Instrument Setup window will be opened.
-
Click the Torch Align
- Perform a horizontal scan
by clicking Torch Scan on the Torch
Align page.
- (When you perform a
torch alignment scan, you should expect to see a scan result similar to a guass curve.)
-
Put sample tubing back into MilliQ water.
Developing a method
Analytical Calibration and Sample Analysis
- Develop a method which includes at least one standard. Standards are added
on the Standards page of the Method Editor.
- From the Worksheet window, click the Sequence tab and do the
following:
a) Ensure the Sample source is set to Manual.
b) Click the Sequence Editor icon and do the
following:
On the Samples and Calibrations page,
- Enter the required number of samples in the Sample count field.
- Select Begin with calibration and Include a blank in calibration.
Check the other boxes as required.
- Click OK. You will be prompted to confirm your selections.
- Click the Analysis tab and do the following:
- In the Analysis Worksheet, select only the Blank and Standard/s to include
in the analysis by clicking in the first column of each row to highlight them.
- Click the green arrow icon to begin the analysis and follow the prompts.
Turn off the ICP
- Rinse the spraychamber by running HNO3 and DI water
through the system for few minutes.
- Extinguish the plasma by select the Plasma Off icon. The pump stops
automatically when the plasma is extinguished.
- Loose the pump tubes by releasing the pressure bars and lift the tubes out
of the grooves.
- Turn off cooler.
- Turn off ICP.
- turn off gas cylinder.
- Exit the ICP Expert software.
- Log off the computer.
Molecular and Nanoscale Analysis Laboratory
Environmental Engineering Science Program