![]()
ENVIRONMENTAL ENGINEERING SCIENCE
Molecular and Nanoscale Analysis Laboratory
BET Surface Area and Pore Volume
Measurement Instrument
Gemini 2375, Micrometrics (Acquired 2001)
Properties as pore size, shape, and chemically active sites that control the reactivity of solids can be evaluated through gas-solid interactions since the structure and reactivity of solid materials determine how their surfaces will interact with gas molecules.
|
Performance of the Autosorb-1 |
|
|
Surface Analysis |
|
|
Nitrogen range |
from 0.05 m2/g to no known upper limit |
|
|
|
|
|
|
|
Pore Analysis |
|
|
Detectable volume limit |
<0.0001cc/g |
|
Pore diameter range |
3.5 to 5000 angstroms with nitrogen |
|
|
|
|
Functional Capabilities |
|
|
Surface analysis |
|
|
Mesopore size distribution |
|
|
Standard micropore methods |
|
|
(A comprehensive range of surface area and pore size methods is available) |
|
Standard
Operating Procedure
(click here to download a pdf version of the information below)
BET Standard Operating Procedures
Index
Preparations for sample analysis
Preparations for sample analysis:
· Fill Liquid Nitrogen Dewar (note: do not tighten cap of dewar tight, this will prevent pressure build up)
· Fill cold trap dewar with liquid nitrogen
o Fill about 2 in from lip of container
· Cold trap dewar is then mounted to autosorb-1.
o Lift and turn cold trap dewar so that bracket (A) is turned to the left and trap cell is inserted into dewar.
o Rotate container clockwise until mounting bracket (A) is even with bracket (B) on sample preparation station.
o Slide bracket (A) into (B).
· Operator must sign in log book
· Open He and N2 cylinder valves and ball valves located after pressure regulators. Confirm cylinder pressure regulators are set to 10 psi.
· Insure the cold trap dewar is half full with liquid nitrogen if not fill according to above directions.
Top of page (Beginning of SOP)
Degassing samples prior to analysis purges impurities.
· Choose appropriate cell size for sample (insert pgs. V-5,V-6 and V-7)
o 6, 9 or 12 mm.
o insert clean fill rod into cell with sample.
· Using a5-place analytical balance, weigh and record the weight of the empty sample cell with the cell holder (a plastic cup with bubble wrap). Tare the weight and then fill the bulb with the recommended weight of the sample. Record the weight of the sample as Initial weight.
· Insert the proper rod into the cell and place cell into the heating mantle
· Attach cell to sample preparation station (for outgas port 1 or 2).
o Remove Knurled retainer ring from sample preparation station (insert pg III-2).
o Assembly contains 3 parts: Knurled retainer ring, plunger and cell adapter (insert pg. VII-9). Remove plunger and use only the retainer ring and cell adapter.
o Choose appropriate size of cell adapter and assemble according to diagram (insert pg. V-8).
· Set temperature (Insert pg. III-2) manually.
o Temperature should be set according to sample specifications.
o 100C allows for water evaporation.
· Pinch heating mantle to open the mouth. Insert cell into mantle. Pinch mantle to close and secure the mantle to cell (insert pg. III-2).
· Turn heat on by lifting the toggle switch to the up position.
· Degassing may take more than 4 hrs.
Top of page (Beginning of SOP)
· On Tool bar click <OPERATION>
· From the drop down menu click <OUTGASSER>
· From pop up menu click on Station 1 or station 2 for the station you are using. If you are using both then click on both.
· Click <LOAD>
· Click OK
· This begins the sample outgas routine.
· Record the start time.
Check sample for proper degassing.
· On the Autosorb go to <OPERATIONS>
· Click on <MENU>
· Click on <OUTGASSER>
· Select the station to test and press <TEST>
· If the Outgasser test is passed in 1 min the sample is ready for analysis. If the Outgas test continues to fail, continue Outgassing and click <OK>. Check the cell ports to make sure the sample is properly loaded.
· When Outgassing is complete, remove sample by going to <OPERATIONS><MENU><OUTGASSER>.
· Select the sample you want to transfer and press <REMOVE>.
· When the LED indicator on the selected port turns orange, the sample is ready for transfer.
· Remove heating mantle and allow sample cell to cool. Once cool, quickly but carefully transfer the sample cell to the test port.
Top of page (Beginning of SOP)
Top of page (Beginning of SOP)
Top of page (Beginning of SOP)